Intermediate precision is one of the most important characteristics for evaluation of precision in assay validation. The current methods for evaluation of within-device precision recommended by the Clinical Laboratory Standard Institute (CLSI) guideline EP5-A2 are based on the point estimator. On the other hand, in addition to point estimators, confidence intervals can provide a range for the within-device precision with a probability statement. Therefore, we suggest a confidence interval approach for assessment of the within-device precision. Furthermore, under the two-stage nested random-effects model recommended by the approved CLSI guideline EP5-A2, in addition to the current Satterthwaite's approximation and the modified large sample (MLS) methods, we apply the technique of generalized pivotal quantities (GPQ) to derive the confidence interval for the within-device precision. The data from the approved CLSI guideline EP5-A2 illustrate the applications of the confidence interval approach and comparison of results between the three methods. Results of a simulation study on the coverage probability and expected length of the three methods are reported. The proposed method of the GPQ-based confidence intervals is also extended to consider the between-laboratories variation for precision assessment.
Date:
2009-09
Relation:
Journal of Biopharmaceutical Statistics. 2009 Sep;19(5):763-778.