    题名: The effect of clockwise and counterclockwise twisting moments on abutment screw loosening
    作者: Yao, KT;Kao, HC;Cheng, CK;Fang, HW;Yip, SW;Hsu, ML
    贡献者: Division of Medical Engineering Research
    摘要: Objectives Abutment screw loosening is one of the most common complications of dental implants. When occlusal loading is applied, bending and twisting moments are counteracted by the implant-abutment connection held with the abutment screw. Restorations in different quarters of the oral cavity might bear clockwise or counterclockwise twisting moments that result from a regular chewing pattern. The aim of this study was to investigate the effect of different directional twisting moments on abutment screws. Methods Twenty 3i implants were divided into four groups of C, R, L, and O. Each assembly consisted of an implant, abutment, and superstructure. For group C, a cyclic load of 10–100 N for 106 cycles was applied centrally and perpendicular to the long axis of the implant, whereas for groups R and L, the same load conditions were applied eccentrically in clockwise and counterclockwise directions, respectively. Group O was left unloaded. The difference between the initial torque and post-load reverse torque was defined as the total torque loss. The data were analyzed by Kruskal–Wallis test. The surface of the abutment hexagon corners were examined with SEM after loading. Results No abutment screw loosening was found after loading. Total torque loss of groups C, R, L, and O were 10.50 ± 0.58, 9.56 ± 1.01, 9.98 ± 1.81, and 9.58 ± 0.94 Ncm, respectively. There were no statistical differences among the four groups. SEM observations showed marked burnishing at the hexagonal corners on the compression sides of the R and L groups. Conclusions Within the limitations of this study, there was little effect of twisting moment direction on the total torque loss of an internal hexagon connection implant system. This could be attributed to the anti-twisting mechanism of the internal hexagon connection.
    日期: 2012-10
    關聯: Clinical Oral Implants Research. 2012 Oct;23(10):1181-1186.
    Link to: http://dx.doi.org/10.1111/j.1600-0501.2011.02282.x
    JIF/Ranking 2023: http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=NHRI&SrcApp=NHRI_IR&KeyISSN=0905-7161&DestApp=IC2JCR
    Cited Times(WOS): https://www.webofscience.com/wos/woscc/full-record/WOS:000308938600009
    Cited Times(Scopus): http://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=84866485762
    显示于类别:[其他] 期刊論文

    文件中的档案:

    档案 描述 大小格式浏览次数
    WIL2012072618.pdf246KbAdobe PDF296检视/开启


    在NHRI中所有的数据项都受到原著作权保护.

    TAIR相关文章

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回馈