    题名: Measurement accuracy of a stand-alone oscillometric central blood pressure monitor: A validation report for microlife watchBP office central
    作者: Cheng, HM;Sung, SH;Shih, YT;Chuang, SY;Yu, WC;Chen, CH
    贡献者: Division of Health Services and Preventive Medicine
    摘要: BACKGROUND The superiority of prognostic value of blood pressure (BP) measured at central aorta (CBP) over conventional brachial BP measured by cuff-based BP monitors has reignited the development of new non-invasive techniques for estimating CBP. The present study validated the accuracy of CBP measured by a newly developed stand-alone CBP monitor. METHODS The CBP monitor provided readings of brachial systolic BP (SBP), brachial diastolic BP (DBP), central SBP, and central pulse pressure (PP). Brachial PP and central DBP were calculated from the relevant readings. The accuracy of the brachial and central SBP, PP, and DBP was validated against the simultaneously recorded invasively measured central aortic SBP, PP, and DBP, according to the invasive standard requirements for the noninvasive brachial BP monitors from the Association for the Advancement of Medical Instrumentation (AAMI) in 85 subjects (255 measurements; age range, 30-93 years). RESULTS The mean differences of cuff BP with reference to the invasively measured central SBP, PP, and DBP were -2.6 +/- 9.0, -8.6 +/- 11.2, and 6.1 +/- 7.0 mm Hg, respectively, with the former two being obviously underestimated at high CBP and overestimated at low CBP. In contrast, the corresponding differences for the central SBP, PP, and DBP measured by the CBP monitor were -0.6 +/- 5.5, -0.4 +/- 7.0, and -0.2 +/- 6.5 mm Hg, respectively, without obvious systematic bias. The distribution of measurement errors for central SBP, PP, and DBP surpassed the AAMI criteria. CONCLUSION Central SBP, PP, and DBP can be measured accurately by a stand-alone automatic BP monitor.
    日期: 2013-01
    關聯: American Journal of Hypertension. 2013 Jan;26(1):42-50.
    Link to: http://dx.doi.org/10.1093/ajh/hps021
    JIF/Ranking 2023: http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=NHRI&SrcApp=NHRI_IR&KeyISSN=0895-7061&DestApp=IC2JCR
    Cited Times(WOS): https://www.webofscience.com/wos/woscc/full-record/WOS:000313972600008
    Cited Times(Scopus): http://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=84876589759
    显示于类别:[莊紹源] 期刊論文

    文件中的档案:

    档案 描述 大小格式浏览次数
    ISI000313972600008.pdf706KbAdobe PDF262检视/开启


    在NHRI中所有的数据项都受到原著作权保护.

    TAIR相关文章

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回馈