    題名: Diagnostic performance of a stand-alone Central Blood Pressure Monitor: Application of Central Blood Pressure in the diagnosis of high blood pressure
    作者: Cheng, HM;Sung, SH;Chuang, SY;Pearson, A;Tufanaru, C;White, S;Yu, WC;Chen, CH
    貢獻者: Division of Health Services and Preventive Medicine
    摘要: Oscillometric central blood pressure (CBP) monitors have emerged as a new technology for blood pressure (BP) measurements. With a newly proposed diagnostic threshold for CBP, we investigated the diagnostic performance of a novel CBP monitor. We recruited a consecutive series of 138 subjects (aged 3093 years) without previous use of antihypertensive agents for simultaneous invasive and noninvasive measurements of BP in a catheterization laboratory. With the cutoff (CBP 130/90mm Hg) for high blood pressure (HBP), the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of the novel CBP monitor were calculated with reference to the measured CBP. In comparison, the diagnostic performance of the conventional cuff BP was also evaluated. The noninvasive CBP for detecting HBP in a sample with a prevalence of 52% showed a sensitivity of 93% (95% confidence interval (CI) 9195), specificity of 95% (95% CI 9497), PPV of 96% (95% CI 9497), and NPV of 93% (95% CI 9095). In contrast, with cuff BP and the traditional HBP criterion (cuff BP 140/90mm Hg), the sensitivity, specificity, PPV, and NPV were 49% (95% CI 4453), 94% (95% CI 9296), 90% (95% CI 8693), and 63% (95% CI 5966), respectively. A stand-alone oscillometric CBP monitor may provide CBP values with acceptable diagnostic accuracy. However, with reference to invasively measured CBP, cuff BP had low sensitivity and NPV, which could render possible management inaccessible to a considerable proportion of HBP patients, who may be identifiable through noninvasive CBP measurements from the CBP monitor.
    日期: 2014-03
    關聯: American Journal of Hypertension. 2014 Mar;27(3):382-391.
    Link to: http://dx.doi.org/10.1093/ajh/hpt282
    JIF/Ranking 2023: http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=NHRI&SrcApp=NHRI_IR&KeyISSN=0895-7061&DestApp=IC2JCR
    Cited Times(WOS): https://www.webofscience.com/wos/woscc/full-record/WOS:000331842300014
    Cited Times(Scopus): http://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=84894228427
    顯示於類別:[莊紹源] 期刊論文

    文件中的檔案:

    檔案 描述 大小格式瀏覽次數
    ISI000331842300014.pdf821KbAdobe PDF424檢視/開啟


    在NHRI中所有的資料項目都受到原著作權保護.

    TAIR相關文章

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回饋