    题名: Determination of the maleic acid in rat urine and serum samples by isotope dilution-liquid chromatography-tandem mass spectrometry with on-line solid phase extraction
    作者: Chen, HC;Wu, C;Wu, KY
    贡献者: National Environmental Health Research Center
    摘要: A rapid and simple on-line solid-phase extraction coupled with isotope dilution-liquid chromatography-tandem mass spectrometry (SPE-ID-LC-MS/MS) method was developed to quantitate maleic acid in serum and urine of Sprague Dawley (SD) rats. The aforementioned biological samples were spiked with 13C2-maleic acid, vigorously vortexed, added with acetonitrile to precipitate proteins, and then injected into the on-line SPE LC-MS/MS system for quantification. Upon validation, this method demonstrated excellent feasibility and sensitivity: Calibration curves for maleic acid in serum and urine display excellent linearity with the coefficient of determination (R2) greater than 0.999, the limits of detection and quantitation (LOD and LOQ) for maleic acid were determined at 0.2 and 0.5 μg L?1, respectively. Additionally, intra-day accuracy for maleic acid in serum and urine samples ranged from 94.0 to 100.2% and 101.3 to 104.4%, respectively. Furthermore, inter-day accuracy ranged from 93.6 to 101.0% and from 102.3 to 111.4% in serum and urine samples, respectively. Intra-day precision %RSD of maleic acid in serum and urine samples was 13.8% or less, whereas the inter-day precision was 6.1% or less. The matrix effects were not found to be statistically significant (p = 0.9145 and p = 0.5378, respectively) based on the calculations of recovery functions. The collected serum and urine samples were analyzed using SPE-ID-LC-MS/MS. Our results reveal trace levels of maleic acid in the control rats; demonstrating that this method is capable of analyzing background levels of contaminants in biofluids with excellent sensitivity and specificity at parts-per-billion levels concentrations in complex matrices.
    日期: 2015-05
    關聯: Talanta. 2015 May;136:9-14.
    Link to: http://dx.doi.org/10.1016/j.talanta.2014.11.021
    JIF/Ranking 2023: http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=NHRI&SrcApp=NHRI_IR&KeyISSN=0039-9140&DestApp=IC2JCR
    Cited Times(WOS): https://www.webofscience.com/wos/woscc/full-record/WOS:000350834700002
    Cited Times(Scopus): http://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=84921527781
    显示于类别:[吳焜裕(2001-2008)] 期刊論文

    文件中的档案:

    档案 描述 大小格式浏览次数
    SDO0039914014009138.pdf1404KbAdobe PDF601检视/开启


    在NHRI中所有的数据项都受到原著作权保护.

    TAIR相关文章

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回馈