    題名: Wearable inverse light-emitting diode sensor for measuring light intensity at specific wavelengths in light therapy
    作者: Tsai, HY;Su, FC;Chou, CH;Lin, YH;Huang, KC;Yang, YJJPD;Kuo, LW;Liao, LD;Yu, HS
    貢獻者: National Institute of Environmental Health Sciences;Institute of Biomedical Engineering and Nanomedicine
    摘要: Optical measurement is conducted by employing components such as a spectrometer or photodiode (PD) to detect light intensity. However, the size of spectrometer is large because of its complex optical path and component. Moreover, a PD usually detects the total light intensity instead of the intensity of light at a specific wavelength. Therefore, a novel inverse light-emitting diode (LED) sensor was developed to measure light intensity at specific wavelengths because of the reverse effect of LEDs. In the experiments conducted in this paper, red, green, and blue LEDs were used in combination as the inverse LED sensor and light source. Moreover, commercial equipment such as a PD, an illuminometer, a power meter, and Actiwatch 2 were simultaneously employed to detect the light intensities. The results revealed that the intensity ratio evaluated using the inverse LED sensor was proportional to that of the light source. This implies that the correlation between the incident light and measured intensities from the proposed inverse LED sensor was stronger than commercial equipment. The relationship between intensity ratio, irradiance distance, and illumination angle was also determined. The least difference between the measured and ideal evaluated intensity ratio was 2.4%. The proposed inverse LED sensor has following advantages: low cost, small size, high-wavelength selectivity, and high-intensity correlation. In a future study, light intensity can be measured using wearable devices comprising the proposed inverse LED sensor, and the presented relationship can be evaluated to investigate the effect between the light intensity and physiological mechanisms or diseases.
    日期: 2019-03-25
    關聯: IEEE Transactions on Instrumentation and Measurement. 2019 Mar 25;68(5):1561-1574.
    Link to: http://dx.doi.org/10.1109/TIM.2019.2899444
    JIF/Ranking 2023: http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=NHRI&SrcApp=NHRI_IR&KeyISSN=0018-9456&DestApp=IC2JCR
    Cited Times(WOS): https://www.webofscience.com/wos/woscc/full-record/WOS:000465231600035
    Cited Times(Scopus): https://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=85064697965
    顯示於類別:[余幸司] 期刊論文
    [郭立威] 期刊論文
    [廖倫德] 期刊論文

    文件中的檔案:

    檔案 大小格式瀏覽次數
    SCP85064697965.pdf3896KbAdobe PDF649檢視/開啟


    在NHRI中所有的資料項目都受到原著作權保護.

    TAIR相關文章

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回饋